2.Cad.No:11/50 06824 ANKARA   Login   Sign Up
PRESSURE GAUGES   THERMOMETERS TEMPERATURE AND PRESSURE SWITCHES   TEST EQUIPMENTS
           
 Home >>Test equipment >>Process Calibrators>>ADT 220
ADT 220   The ADT 220 is a highly integrated loop calibrator featuring several patented technologies, the calibrator is an ultra-compact, rugged, and best of all, easy to use hand-held device that will source, simulate and measure loop current, mV/V, loop integrity, and also perform switch testing. Its smart phone-like menu and interface makes it simple and easy to use.
  •       SUMMARY        
  •      SPECIFICATIONS    
  •       ORDER CODE        
  •          PRICE               
  •     DOWNLOADS     
• Measuring and Sourcing loop current, mV, and V,
• Measuring and sourcing simultaneously,
• Loop integrity check capability,
• 24V loop supply with simultaneous current measurement,
• High-capacity storage with up to twenty thousand records (Data logging),
• At the same time two readings, 3.5 inch TFT color screen,
• Smartphone-like menu and interface make the operation simpler and easier,
• 3 years warranty,
• NIST Traceable Calibration with data.
 
 
 
 
 
 
  Operation temperature -10°C to 50°C
  Humidity <90%
  Certification CE
  Display 3.5 inch TFT color screen
  Data statistics and analysis Display max, min, peak-to-peak, average and percentage simultaneously
  Data logger High-capacity storage with up to twenty thousand records
  Snapshot Save and manage up to 100 snapshots
  Electrical connection (Ø) 4mm socket and mini-jack TC socket
  Communication RS232; standard RS232-DB9
  Dimensions 100mm x 192mm x 52mm
  Weight 700 gr.
  Supply Polymer Li-ion rechargeable battery or 10V DC adaptor
  Battery lifetime 15 hours uninterrupted use
Battery life will be reduced when 24V is applied.
  Battery filling 110V/220V external adapter (included)
  Warranty 3 years
  Electrical properties
  Measurement  Range Resolution Accuracy
Voltage DC -300.000 to 300.000 mV
-60.0000 to 60.0000 V
1µV 0.1 mV 0.01%RD+15 µV
0.01%RD+3 mV
Current DC -30.0000 to 30.0000 mA 0.1 µA 0.01%RD+1.5 µA
Loop integrity test 0 to 2000.0 ohm 0.1 Ohm 0.02%RD+0.2 Ohm
Switch test In case the switch comes with voltage, the voltage range: 3 to 24V
  Source
Voltage DC -10.00 to 200.00 mV
0 to 12.000 V
0.01mV
1mV
0.02%RD+10.5 µV
0.02%RD+0.6 mV
Current DC 0 to 22.000 mA 1µA 0.02%RD+1.1 µA
DC24V N/A N/A 0.5V